ISO 15883-1:2024

مواصفة قياسية دولية   الإصدار الحالي · اعتمدت بتاريخ ٠٤ أكتوبر ٢٠٢٤

Washer-disinfectors — Part 1: General requirements, terms and definitions and tests

ملفات الوثيقة ISO 15883-1:2024

الإنجليزية 73 صفحات
الإصدار الحالي
OMR 93.51

مجال الوثيقة ISO 15883-1:2024

This document specifies general performance requirements for washer-disinfectors (WD) and washer-disinfector accessories that are intended to be used for cleaning and disinfection of reusable medical devices. It specifies performance requirements for cleaning and disinfection as well as for the accessories that can be required to achieve the necessary performance. The methods and instrumentation required for validation, routine control and monitoring and requalification, periodically and after essential repairs, are also specified.

NOTE 1        The requirements can be applied to WD intended for use with other articles used in the context of medical, dental, pharmaceutical and veterinary practice.

The requirements for WD intended to process specific loads are specified in ISO 15883-2, ISO 15883-3, ISO 15883-4, ISO 15883-6 and ISO 15883-7. For WD intended to process loads of two or more different types, the requirements of the applicable parts of ISO 15883-2, ISO 15883-3, ISO 15883-4, ISO 15883-6 and ISO 15883-7 apply.

This document does not specify requirements intended for machines for use for laundry or general catering purposes.

This document does not include requirements for machines which are intended to sterilize the load, or which are designated as “sterilizers” and addressed in other standards.

The specified performance requirements of this document do not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.

NOTE 2        Chemicals in some cleaning agents and disinfectants can react with prion protein in a manner that can inhibit its removal or inactivation.  If the presence of prion protein is considered a possibility, then this can influence the choice of cleaning agent and disinfectant.

NOTE 3        This document can be used by prospective purchasers and manufacturers as the basis of agreement on the specification of a WD. The test methods for demonstration of conformity with the requirements of this document can also be employed by users to demonstrate continued conformity of the installed WD throughout its service life. Guidance on a routine test programme is given in Annex A.

الأكثر مبيعاً

GSO 150-2:2013
 
مواصفة قياسية خليجية
فترات صلاحية المنتجات الغذائية - الجزء الثاني : فترات الصلاحية الاختيارية
GSO 9:2022
 
لائحة فنية خليجية
بطاقات المواد الغذائية المعبأة
GSO 2055-1:2015
 
لائحة فنية خليجية
منتجات الحلال – الجزء 1: الاشتراطات العامة للأغذية الحلال
GSO 150-1:2013
 
لائحة فنية خليجية
فترات صلاحية المنتجات الغذائية - الجزء الأول : فترات الصلاحية الإلزامية

اعتمدت مؤخراً

ISO 21815-4:2026
 
مواصفة قياسية دولية
Earth-moving machinery — Collision warning and avoidance — Part 4: Risk area and risk level for swing and rotation motion
ISO 21611:2026
 
مواصفة قياسية دولية
Dentistry — Vocabulary for source conclusion for human identification by dental evidence
ISO 4576:2026
 
مواصفة قياسية دولية
Plastics — Polymer dispersions — Determination of sieve residue
ISO 8553:2026
 
مواصفة قياسية دولية
Milk — Enumeration of microorganisms — Plate-loop technique at 30 °C